Measuring Equipment » Drop Volume Tensiometer TVT 2

Technical data


Measuring range mN/m 0.1...100
Resolution    
- Resolution µm ± 0.1
- Volume µl ± 0.01
- Surface/interfacial tension mN/m ± 0.01
- Drop formation time s ± 0.1
Reproducibility of individual drops for pure liquids (mechanical tolerances)    
- Stroke µm < 2.5
- Volume (based on syringe volume) ‰ µl 0.07
- Surface tension mN/m ± 0.08 x syringe volume [ml] (1)
- Interfacial tension   The above value must be multiplied by δ (density difference)
- Drop formation time for t <100 s (2) s ± 0.1 ... 0.5
Reproducibility of mean value over 5 drops    
- Surface tension (depending on syringe and needle type) mN/m ± 0.01 ... 0.05
Absolute accuracy   approx. 0.5% of end value of surface tension
Drop times s/µl 0.04 (at 5 ml) 170 (at up to 0.25 ml)
Speed control < 1 %  
Temperature range °C 5...90 (with special thermostating block)
Interfaces   RS 232
Dimensions of TVT 2 measuring desk (WxDxH) mm 220 x 240 x 555
Dimensions of TVT 2 control unit (WxDxH) mm 340 x 270 x 105
Weight of TVT 2 measuring desk kg 8.0
Weight of TVT 2 control unit kg 4.2
Power consumption W 0.1
Power supply V; Hz 80...230; 50/60

(1) this means, for instance:

Syringe volume [ml] 0.25 0.5 1 2.5 5

Ds[mN/m] ± 0.02 ± 0.04 ± 0.08 ± 0.16 ± 0.32

(2) depending on drop size and discharge speed

Standard accessories

  • Software
  • RS 232 cable
  • Mains cable
  • Connecting cable for measurement desk/control unit
  • Light barrier cable (transmitter and receiver)
  • Syringe 2.5 ml
  • Standard needle SK1
  • Cell
  • Thermostating block(60 °C or 90 °C model)
  • Allen key
  • Cell handling tool

Optional accessories

  • Syringe with various volumes
  • Thermostating block (max. 60 °C or alternatively 90 °C)
  • Spare screw cap for needles
  • Temperature probe
  • Cell handling tool
  • Adapter
  • RS 232 cable
  • Steel needles
  • Reverse needles
  • Glass needles
  • Cell (heat-proof glass)
  • Reverse measuring set

Metrological standards

  • ISO 9101
  • ASTM D2285